Device Not Okay

Error 009-032 for USB Devices using MagIC Net, Tiamo, and StabNet 

Technology: Ion Chromatography – MagIC Net, Titration – Tiamo, Stability – StabNet  

General  

This document details how to troubleshoot USB instrument connectivity, including how to resolve error 009-032 – “Device not ok,” which is a common error received for Metrohm software Tiamo, MagIC Net, and StabNet. 

 

A screenshot of a computer error

Description automatically generated 

Procedure 

  • A. Confirm Status of Device in Configuration  

  • Go to the Configuration tab and locate the Devices sub-window, which is commonly found in the top left quadrant.  
  • Verify the instrument’s Status.  
    • If the Status bar is yellow and is not OK, the instrument is still in the initialization phase. Wait a few minutes and watch the status move to either OK or Not OK.  
    • If the status is listed as OK, proceed as usual. Contact Tech Support for further assistance if the Device Not Okay 009-032 error message is still received with Status OK.  
  • If the status is in red and listed as not OK, continue to Part B.  

A screenshot of a computer

Description automatically generated 

  •  
  • B. Verify Computer Recognizes the Instrument: 

  • On the computer, open Device Manager: Press the Windows Key    + X, then select Device Manager. 
  • Scroll down to the USB Metrohm folder, then double-click to expand.  
  • If the instrument is listed in Device Manager, continue to part C: Establish Connection.  
  • If the instrument isn’t listed in Device Manager, skip to part D. 
 

A screenshot of a computer

Description automatically generated 

  • C. Establish Connection 

  • Disconnect the instrument from the computer.
  • In software > Configuration > Device sub window, right-click on the instrument and select Delete.
  • Close the software and reconnect the instrument to the computer. 
  • Verify that the instrument is listed again in Device Manager > USB Metrohm folder. 
  • When opening the software, A window should appear as 009-108. Save the device, then click Yes > OK. 
  • Click Configuration – wait for the status OK, then check the method. 
  • If the instrument is listed in Device Manager and window 009-108 doesn’t generate:  
    • Remove the main power from the instrument, close the software, and disconnect any auxiliary components, such as 800 Dosinos and/or stirrers plugged into the MSB ports. 
    • Be sure to note what ports the components are connected to. 
  • Return power to the instrument and open the software with disconnected components.  
  • In the configuration, if the instrument’s status is OK, reconnect one MSB device at a time to determine the source of the error.  
    • When connecting or reconnecting any MSB device, the power must be removed to avoid problems from “hot swapping”.
  • Contact Tech Support if the device status is not OK.
 
  • D. Troubleshoot - If the USB Metrohm Folder Isn’t Available:  

  • Verify the instrument has power by checking the main power cable on the instrument to the power outlet or uninterrupted power supply.  
  • For a titrator or autosampler, a controller cable 62151000 is connected to the instrument’s controller port on the instrument's back panel.  
 

  • Device not Okay.png

  • When removing or connecting a new controller cable –power should be removed from the instrument.
  • Disconnect the main power supply.  
  • Remove the cable and inspect that eight pins are present, not corroded or bent. 
  • Ensure that the controller port isn’t damaged or has a broken pin inside. 
  • Confirm this connection is properly secured in the back panel of the instrument.
  • Verify the USB connection is connected to the PC’s USB port.  
  • Try a different USB port and different controller cable.  
  • For an IC and Rancimat, a USB A to B cable 62151020 connects from the instrument to the computer. This is a standard printer-to-PC cable.  
  • If the instrument still isn’t recognized by the computer in the USB Metrohm folder, then contact Metrohm Tech Support for additional assistance. 
 
  • If there is an unknown device listed, as shown below. This confirms that the device driver is not installed properly. Two options are available.  

A screenshot of a computer

Description automatically generated 

1. Reinstall the Device Driver: 

        • Right click the device and choose properties. 
        • Choose the driver tab. 
        • Click update driver.
        • Choose browse my computer for driver software.
        • Search the folder C:\Program Files\Metrohm\metr_770. 
        • Select the option include subfolders. 
        • Press Next.
 
  • 2. Uninstall and Reinstall the Device: 

        • Right-click the device and choose uninstall device. 
        • Disconnect the USB cable from the PC.
        • Plug the USB cable back in slowly. It may also be beneficial to try a different USB port. 
 

If the instrument still isn’t recognized by the computer in the USB Metrohm folder, contact Metrohm Tech Support for additional assistance.

Prevention 

Ensure the instrument is connected to an uninterrupted power supply to prevent power outage related issues.  

Additional Support 

If you encounter any issues or need additional help, please submit a support ticket through Metrohm Technical Support. 

For additional training on any of these topics, explore our available Metrohm Ion Chromatography Training courses or learn more about our Metrohm Custom Training options. 