874 KF Oven Gas Flow Error

Tiamo error 009-141, 874 error 13-4-1-255 

Technology: Titration

How do you know you have this problem?

Tiamo software generates the following error:

A screenshot of a computer error message

Description automatically generated

What can I try to resolve it?

The majority of the time, a gas flow error is caused by blockage. 

Follow these steps to troubleshoot the gas flow error:

  • Ensure all tubing connections are tight.
  • Ensure the gas tubing in the vessel is not crimped.
  • If using pressurized air or gas, ensure the following:
    • Regulator is set to < 4psi.
    • The tank is not empty.
    • The valve on the cylinder is open.
  • Inspect needle holder (6.2049.040 or 6.2049.050) for damage. 
  • Clean or replace injection needle (6.2816.070) and outlet needle (6.2816.080).
    • Needles can be flushed with methanol or sonicated for faster cleaning.

A diagram of a machine

Description automatically generated

 

  • Clean or replace the transfer tube
    • Tube can be flushed with methanol.
  1.  

A diagram of a wire connected to a cable

Description automatically generated

Shape

Here is a video to walk you through the process: HOW TO troubleshoot gas flow issues on a Metrohm 874 Karl Fischer Oven system

  • Evaluate impact of molecular sieve by connecting tubing for ‘to’ and ‘from’ of the tower to bypass the molecular sieve bottles.
    • Disconnect ‘from’ tubing from the tower (see below).
    • Disconnect ‘to’ tubing from the cap of molecular sieve bottle (see below).
    • Connect this end of the ‘to’ tubing to the ‘from’ port on the tower.
    • If error no longer occurs with sieve bypassed, replace molecular sieve.
  1.  

What are some ways to prevent the problem?

  • Use Metrohm KF oven vials and septum caps.
  • Regularly inspect and clean needles / transfer tube.
  • Regularly inspect, regenerate or replace molecular sieve.
    • The molecular sieve can be regenerated at 300 °C in the drying oven. The regeneration time is at least 24 hours. The molecular sieve is then placed in a desiccator for cooling and afterwards stored in a glass bottle with an airtight seal.