LL-Ag/AgCl (Gel) Reference System - Regenerating a Dried Out Electrode

Introduction

If an LL-Ag/AgCl (gel) reference electrode has dried out, it cannot be regenerated. Unlike refillable reference systems, both the maintenance-free 6.0728.x40 reference system and the gel-filled electrolyte vessel must be replaced if they have dried out. Attempting to restore either component will not return the electrode to proper operating condition.

This article explains how to identify a dried-out LL-Ag/AgCl (gel) reference electrode and outlines the required replacement procedure. If the maintenance-free reference system has dried out, the reference system must be replaced. Likewise, if the KCl gel in the electrolyte vessel has dried out, the electrolyte vessel must also be replaced. Replacing the affected components ensures a stable reference potential and reliable electrochemical measurement performance.

Video

Topics Covered

  • Neither the reference system nor the electrolyte vessel can be regenerated.
  • Reference system (0:06)
    • Once the 6.0728.x40 reference system has dried out, it cannot be regenerated. The electrode must be replaced.
  • Electrolyte vessel (0:18)
    • Dried-out KCl gel cannot be removed from the electrolyte container. The electrolyte container must be replaced.

Additional Support

If you encounter any issues or need additional help, please submit a support ticket through Metrohm Technical Support. 

For more detailed training on any of these topics, explore our Metrohm Custom Training options. 

Was this article helpful?