884 Professional VA with Ultra Trace-RDE

Introduction

The Ultra Trace Rotating Disk Electrode (Ultra Trace RDE) is designed for voltammetric applications that require high-sensitivity trace analysis with a solid-state working electrode. To ensure reliable performance, the Ultra Trace RDE must be used with dedicated platinum-free accessories. Instead of a platinum auxiliary electrode, a glassy carbon rod is used. Before beginning, gather the required components, including the Ultra Trace RDE drive shaft and electrode tip, reference electrode, glassy carbon auxiliary electrode, the RDE measuring head, and the appropriate electrode cables.

During setup, the Ultra Trace RDE is installed in the RDE measuring head, followed by the installation of the reference and auxiliary electrodes. Each electrode is then connected to the appropriate instrument input. Before use, the electrode surface should be regenerated using the Ultra Trace polishing set (6.2802.020) by rotating the electrode at 2000 rpm while polishing with light pressure. The electrode should then be conditioned according to the application instructions to regenerate and activate the electrode surface. Following these preparation steps helps ensure accurate, reproducible voltammetric measurements.

Video

Topics Covered

  • The use of the Ultra Trace RDE requires special accessories. These must be platinum-free and should only be used with this type of electrode. A glassy-carbon rod is used as the auxiliary electrode instead of platinum.
  • The working electrode consists of a drive shaft and an electrode tip.
  • The electrode surface of the Ultra Trace RDE can be regenerated using the special polishing set 6.2802.020.
  • To do this, insert the electrode into the RDE measuring head, rotate it at 2000 rpm, and polish with light pressure.
  • Next, the reference and auxiliary electrodes are added.
  • Connect the electrode cables to the corresponding electrodes:
    WE – Ultra Trace-RDE
    RE – Reference electrode
    AE ​​– Auxiliary electrode
  • Solid-state electrodes must be conditioned to regenerate and activate the electrode surface.
  • Refer to the application instructions for electrode preparation information.

Additional Support

If you encounter any issues or need additional help, please submit a support ticket through Metrohm Technical Support. 

For more detailed training on any of these topics, explore our Metrohm Custom Training options. 

 

Was this article helpful?