Create an Export Template (.csv) in Tiamo

Create an Export Template (.csv) - tiamo 

Guidance for creating a .csv export template. 

Instructions for Template Creation: 

  1. Launch tiamo software.
  2. Click Database tab,
  3. Click Tools Templates Export templates…
  4. Click ‘New’. 
  5. Create a template with the following properties: 
  • Name: CSV (or similar). 
  • Target directory: location on hard drive or network drive (click the button to right of field to search for a folder – click “New Directory” to create a new folder). 
  • File type: *.csv (Comma Separated). 
  • File name: 

    i. Determination ID – automatic unique ID assigned by tiamo.

    ii. Sample identification – file name based on user-entered ID.

    iii. Request on each export. 

    iv. Fixed file name (append data) – file name of your choice (exported data appended to same file). 

6. Click ‘Options’ 

  • Field separator: select the comma. 
  • Record separator: select CR/LF. 
  • Header: Check this if you want the ‘displayed name’ as defined in the ‘Select fields’ menu (see step 7 below) at the top of each column in the export file.

7. Click ‘Select fields’ 

  • Highlight fields from the left panel and click the button in the middle of the window to move to right panel – to be included in export file.
  • Highlight fields from the right panel and click the << button in the middle of the window to move to the right panel - to be removed from the export file. 
  • You can reorder the fields by highlighting the field name and using the up/down arrows on the far right of the window. 

8. Click 'OK' to store new template. 

9. Click 'Close' to exit the template window. 

How to Manually Export

  1. Click Database tab. 
  2. If current Database is not already open, click File > Open and select appropriate database. 
  3. Select all determinations you wish to export in the Determination Overview. 
  4. Click Determinations > Export. 

  5. Select your newly created Export template. 
  6. The files will be exported to the 'Target Directory' as programmed in the template. 

Set up Automatic Export in Method

To automatically export determination(s) after titration is completed, add an EXPORT command into your tiamo method. 

  1. Click Method tab. 
  2. Click File > Open and select method in which you would like to add automatic export. 
  3. Click Insert > New command. 

  4. Double click on 'Results' folder to open. 
  5. Select 'Export'.

  6. Click 'OK' - you will see the new EXPORT command in the Method. 
  7. Click and hold the EXPORT command, drag and place after the CALC command. 

  8. Double click on the EXPORT command to open the Properties of the command. 
  9. Select your newly created export template. 
  10. Click 'OK'. 
  11. Click File > Save to save the updated method. 